Full-text resources of PSJD and other databases are now available in the new Library of Science.
Visit https://bibliotekanauki.pl

PL EN


Preferences help
enabled [disable] Abstract
Number of results
2011 | 17 | 3 | 153-158

Article title

Application of recombination chambers for determination of neutron ambient dose equivalent at the door to the maze of linear accelerator vault

Content

Title variants

Languages of publication

EN

Abstracts

EN
Recombination chambers and 2202D Studsvik-Alnor reference remmeter were used for determination of ambient dose equivalent, H*(10) and its neutron and gamma components at the door to the maze of 15 MV medical accelerator. The measurements were performed at opened and partly closed door. The results showed that the ambient dose equivalent rate was 76.2 μSv h-1 at the opened door and only 2.7 μSv h-1 at slightly opened door, 2.5 m from the entrance to the maze.

Publisher

Year

Volume

17

Issue

3

Pages

153-158

Physical description

Dates

published
1 - 1 - 2011
online
20 - 3 - 2012

Contributors

  • Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, Sw. A. Boboli 8, Warsaw, Poland
  • Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, Sw. A. Boboli 8, Warsaw, Poland
  • National Centre for Nuclear Research, Otwock-Świerk, Poland
author
  • National Centre for Nuclear Research, Otwock-Świerk, Poland

References

  • Golnik N. Recombination methods in the dosimetry of mixed radiation. Swierk (PL): Institute of Atomic Energy, Report No IAE -20/A.
  • Golnik N., Tulik P., Bulski W. RADEM - recombination ambient dose equivalent meter for neutron dosimetry around medical accelerators. IFMBE Proceedings. 2009; 25/3: 484-486.[Crossref]
  • Králik M, Turek K. Characterisation of neutron fields around high-energy X-ray radiotherapy machines. Radiat. Prot. Dosim. 2004; 110(1-4): 503-507.
  • Thomas DJ, Bardell AG, Macaulay EM. Characterisation of a gold foil-based Bonner sphere set and measurements of neutron spectra at a medical accelerator. Nucl Instr. Meth. in Phys. Res A. 2002; 476: 31-35.
  • Zanini A, Durisi E, Fasolo F, Visca L, Ongaro C, Nastasi U, et al. Neutron spectra in a tissue equivalent phantom during photon radiotherapy treatment by Linacs. Radiat. Prot. Dosim. 2004; 110(1-4): 157-160.
  • Zielczynski M, Golnik N. Recombination index of radiation quality - measuring and applications. Radiat. Prot. Dosim. 1994; 52(1-4): 419-422.