EN
Lead (+2) was selectively adsorbed on a solid phase extraction (SPE) gel (molecular recognition technology, MRT), quantitatively extracted, and spectrophotometrically determined as the Pb(II)-PAR (4-(2-pyridylazo)-resorcinol) complex. The linear range was 0.01 to 0.75 mg L−1 and the detection limit was 6.4 µg L−1. The MRT-SPE allows selective Pb(II) extraction from complex ion-rich matrices, which is difficult with other techniques. Interference from common matrix ions such as Fe2+, Ni2+, Cu2+ or Co2+ is minimized. [...]