Full-text resources of PSJD and other databases are now available in the new Library of Science.
Visit https://bibliotekanauki.pl
Preferences help
enabled [disable] Abstract
Number of results

Results found: 1

Number of results on page
first rewind previous Page / 1 next fast forward last

Search results

help Sort By:

help Limit search:
first rewind previous Page / 1 next fast forward last
EN
The object of this work was to investigate structure, electrochemical behavior and semiconducting properties of the TiO₂ oxide layer on the Ti-15Mo implant alloy surface in normal and inflammatory conditions of physiological saline solution. X-ray photoelectron spectroscopy measurements confirm the presence of the oxide layer on the Ti-15Mo alloy surface. Electrochemical studies indicate excellent corrosion resistance of Ti-15Mo alloy in physiological saline solution. It was found that the investigated material under normal and inflammatory conditions behave like an insulator and n-type semiconductor, respectively.
first rewind previous Page / 1 next fast forward last
JavaScript is turned off in your web browser. Turn it on to take full advantage of this site, then refresh the page.