Full-text resources of PSJD and other databases are now available in the new Library of Science.
Visit https://bibliotekanauki.pl
Preferences help
enabled [disable] Abstract
Number of results

Results found: 1

Number of results on page
first rewind previous Page / 1 next fast forward last

Search results

help Sort By:

help Limit search:
first rewind previous Page / 1 next fast forward last
EN
A columnar mini phantom is designed as recommended by ESTRO to measure the Head Scatter Factor (Sc) for 6 MV beam of two linear accelerators. The measurement of Sc at different orientations of the chamber, parallel and perpendicular at 1.5 cm depth predicts the deviation of 2.05% and 1.9% for Elekta and Siemens linear accelerators respectively. The measurement of Sc at 1.5 cm is higher compared to 10 cm depth for both the linear accelerators suggesting the electron contamination at 1.5 cm depth. The effect of wedges on Sc yields a significant contribution of 3.5% and 5% for Siemens and Elekta linear accelerators respectively. The collimator exchange effect reveals the opening of upper jaw increases the Sc irrespective of the linear accelerator. The result emphasizes the need of Sc measurement at 10 cm. The presence of wedge influences the Sc value and the SSD has no influence on Sc. The measured Sc values are in good agreement with the published data.
first rewind previous Page / 1 next fast forward last
JavaScript is turned off in your web browser. Turn it on to take full advantage of this site, then refresh the page.