The limitation of DEXA analysis for bone mass determination in mice

Content
Title variants
Languages of publication
EN
Abstracts
EN
An increase in femoral and tibio/fibular bone mass following periosteal membrane stimulation by Moloney sarcoma virus inoculation into thigh muscles of mice was measured in situ on formalin fixed excised hind limbs using a Hologic 4500A Fan Beam X-ray bone densitometer adapted for small bone samples. These results were verified by measurements of constant dry bone mass of the same bones liberated from soft limb tissues by NaOH hydrolysis. There was no consistent data correlation found between the DEXA scan and dry bone mass evaluations. It is concluded that the sensitivity of the DEXA measurement is unsuitable when assessing very small bone samples, weighing merely 20-30 mg.
Keywords
References
Document Type
ARTICLE
Publication order reference
K.H. Wlodarski, Chair of Histology and Embryology, Biostructure Center, Warsaw Medical Academy, 5 Cha?ubinskiego Str., 02-004 Warsaw, Poland