Content
Title variants
Languages of publication
EN
Abstracts
EN
In this study, a dosimetric study was observed by using Cr-51 gamma radioisotope on liver scintigraphy. In this context, the source organ is determined as liver and it is aimed the evaluation of measuring doses from source organ to selected three critical organs (thyroid, gonad, and brain). An antropomorphic torso phantom has been preferred in the experiments. The measurements of critical organs doses were examined for chronic effects. Moreover, the chronic effect of Cr-51 radioisotope on critical organs was determined as cumulative dose. Thermoluminescent dosimeters were used for chronic dose assessment. In the experiments, lithium fluoride (LiF) (TLD-100) detectors were used and they were provided from Turkish Atomic Energy Authority - Cekmece Nuclear Research and Training Center and calibrated in there. Within this study, the evaluation of the cumulative doses by using Cr-51 for three critical organs was observed with an original setup by using anthropomorphic torso phantom experimentally.
Keywords
Year
Volume
132
Issue
3
Pages
476-478
Physical description
Dates
published
2017-09
References
  • [1] The Free Dictionary, Medical Dictionary, scintigraphy http://medical-dictionary.thefreedictionary.com/scintigraphy
  • [2] ACR-SNM-SPR Practice Guideline for the Performance of Liver and Spleen Scintigraphy http://snmmi.files.cms-plus.com/docs/Liver_Spleen_Scintigraphy_1382731942520_6.pdf
  • [3] T. Kashiwagi, Nuclear Gastroenterology, Hyogo College of Medicine, Hyogo (Japan) 2002
  • [4] Medicine Online, Liver Scintigraphy - Liver-Spleen Scan http://medicineonline.com/articles/L/2/Liver-Scintigraphy/Liver-Spleen-Scan.html
  • [5] B. Buyuk, A.B. Tugrul, Ann. Nucl. Energy 71, 46 (2014), doi: 10.1016/j.anucene.2014.03.026
  • [6] O. Maranci, M.Sc. Thesis, Istanbul Technical University, Energy Institute, Istanbul 2016 http://enerji.itu.edu.tr/egitim/tezler/enerji-enstitusu-tezleri
  • [7] Pacific Northwest X-ray Inc., Accessories, Phantoms Radiology, Torso, PET, 2015 http://www.pnwx.com/Accessories/Phantoms/Radiology/PET/Torso/
  • [8] J.R. Greening, in: Fundamentals of Radiation Dosimetry, Medical Physics Handbooks, Vol. 15, 2nd ed., Adam Hilger, Bristol 1981
  • [9] D. Yaşar, A.B. Tuğrul, Radiat. Measur. 37, 113 (2003), doi: 10.1016/S1350-4487(02)00182-8
  • [10] D. Yaşar, A.B. Tuğrul, Appl. Radiat. Isotopes 62, 405 (2005), doi: 10.1016/j.apradiso.2004.08.001
  • [11] H. Hasabelrasoul, Estimation of uncertainty in TLD calibration http://iaea.org/inis/collection/NCLCollectionStore/_Public/46/040/46040700.pdf
  • [12] G. Yilmaz, A.B. Tugrul, M. Demir, D. Yasar, B. Demir, Acta. Phys. Pol. A 130, 90 (2016), doi: 10.12693/APhysPolA.130.90
  • [13] B. Buyuk, A.B. Tugrul, M. Cengiz, O. Yucel, G. Goller, F.C. Sahin, Acta. Phys. Pol. A 128, B-132 (2015), doi: 10.12693/APhysPolA.128.B-132
  • [14] İ. Akkurt, H.O. Tekin, A. Mesbahi, Acta. Phys. Pol. A 128, B-332 (2015), doi: 10.12693/APhysPolA.128.B-332
  • [15] O. Gurler, U. Akar Tarim, Acta. Phys. Pol. A 130, 236 (2016), doi: 10.12693/APhysPolA.130.236
  • [16] M. Cengiz, S.C. Ozer, B. Buyuk, A.B. Tugrul, O. Yucel, G. Goller, F.C. Sahin, Acta. Phys. Pol. A 128, B-187 (2015), doi: 10.12693/APhysPolA.128.B-187
Document Type
Publication order reference
YADDA identifier
bwmeta1.element.bwnjournal-article-appv132n3p019kz