Monte Carlo Analysis of the Influence of Different Packaging on MOSFET Energy Response to X-rays and Gamma Radiation

Content

Title variants

Languages of publication

EN

Abstracts

EN
Radiation sensing MOSFETs have found numerous applications as detectors or device components in radiation fields used in nuclear industry, medical applications and space research. Monte Carlo simulations of MOSFET energy response to X-ray and gamma radiation for different packaging were performed. The photon transport Monte Carlo software FOTELP-2K10 has been adapted to obtain the energy deposited in MOSFET structure with microscopic dimensions. In this work the ratio between values of total energy deposited in the sensitive volume (thick SiO_2 layer) for cases of MOSFET structure with and without package lid is presented. For this purpose is defined the shielding energy dependence factor (SDEF), and gave its value for kovar and Ti-24Al-11Nb and Ti-13Nb-13Zr alloys as lid materials.

Keywords

EN

Year

Volume

122

Issue

4

Pages

655-658

Physical description

Dates

published
2012-10
received
2011-09-05
(unknown)
2012-07-03

Contributors

author
  • Vinča Institute of Nuclear Sciences, Belgrade, Serbia
author
  • Vinča Institute of Nuclear Sciences, Belgrade, Serbia
  • Vinča Institute of Nuclear Sciences, Belgrade, Serbia
author
  • Institute for Testing Materials-IMS, Belgrade, Serbia
author
  • Faculty of Technology and Metallurgy, University of Belgrade, Belgrade, Serbia

References

  • [1] A. Holmes-Siedle, Nucl. Instrum. Methods 121, 169 (1974)
  • [2] A.B. Rosenfeld, M.I. Reinhard, D. Marinaro, P. Ihnat, G. Taylor, L. Peak, N. Freeman, D. Alexiev, M. Lerch, IEEE Trans. Nuclear Sci. 46, 1766 (1999)
  • [3] G.J. Brucker, S. Kronenberg, F. Gentner, IEEE Trans. Nucl. Sci. 42, 33 (1995)
  • [4] Z. Savic, M. Kovačević, M. Petrović, Radiat. Protect. Dosim. 64, 205 (1996)
  • [5] T. Kron, L. Duggan, A. Rosenfeld, T. Smith, M. Butson, G. Kaplan, S. Howlet, K. Hyodo, Phys. Med. Biol. 43, 3235 (1998)
  • [6] M.A.R. Othman, D.L. Cutajar, N. Hardcastle, S. Guatelli, A.B. Rosenfeld, Radiat. Prot. Dosim. 141, 10 (2010)
  • [7] FOTELP-2K6, Photons, Electrons and Positrons Transport in 3D by Monte Carlo Techniques, IAEA-1388, http://www.nea.fr/tools/abstract/detail/iaea1388
  • [8] FOTELP-2K10, http://www.vinca.rs/%7Erasa/doc/FOTPEN10.PDF
  • [9] S.J. Stanković, R.D. Ilić, P. Osmokrović, B. Lončar, A. Vasić, IEEE Trans. Plasma Sci. 34, 1715 (2006)
  • [10] E.A. Siegbahn, E. Bräuer-Krisch, A. Bravin, H. Nettelbeck, M.L.F. Lerch, A.B. Rosenfeld, Med. Phys. 36, 1128 (2009)
  • [11] S.J. Stanković, R.D. Ilić, M. Petrović, B. Lončar, A. Vasić, Mater. Sci. Forum 518, 361 (2006)
  • [12] S.J. Stanković, M. Petrović, M. Kovačević, A. Vasić, P. Osmokrović, B. Lončar, Mater. Sci. Forum 555, 141 (2007)
  • [13] S.J. Stanković, R.D. Ilic, D.M. Davidovic, M. Petrovic, Acta Phys. Pol. A 115, 816 (2009)
  • [14] S.J. Stanković, R.D. Ilic, K.S. Jankovic, A Vasic-Milovanovic, B. Loncar, Acta Phys. Pol. A 120, 252 (2011)
  • [15] XCOM: Photon Cross Sections Database, 2010, http://physics.nist.gov/PhysRefData/Xcom/html/xcom1.html

Document Type

Publication order reference

Identifiers

YADDA identifier

bwmeta1.element.bwnjournal-article-appv122n403kz
JavaScript is turned off in your web browser. Turn it on to take full advantage of this site, then refresh the page.