PL EN


Preferences help
enabled [disable] Abstract
Number of results
2012 | 121 | 1 | 131-134
Article title

Measurement of Wear in Orthopedic Prosthesis

Content
Title variants
Languages of publication
EN
Abstracts
EN
Wear in orthopedic prosthesis has been the main limiting factor of total joint replacement's service life. Wear analysis of total joint replacements is essential for determining failure mechanism, prediction of wear and longevity of implants. So it would be possible to improve design, material and manufacturing quality and service life of these devices. Radiographic, gravimetric, volumetric and optical techniques are current methods for measuring of wear in retrieved implants. Gravimetric method is the standardized method for quantifying wear volumes of total joint prosthesis. Although this method is effective for determining experimental wear volume in simulated conditions, it is not suitable for assessing the clinically retrieved prosthesis for which there is no pre-wear data available. Each of the wear measuring methods has advantages and limitations. Researches have been going on for developing effective methods. In this paper the techniques that are currently used for evaluating wear of orthopedic prosthesis like gravimetry, coordinate measuring machine, micro computed tomography, and digital photogrammetry are explained. The literature works and new trends in metrological assessment of wear are reviewed.
Keywords
EN
Publisher

Year
Volume
121
Issue
1
Pages
131-134
Physical description
Dates
published
2012-01
Contributors
author
  • Department of Mechanical Engineering, Yildiz Technical University, Istanbul, Turkey
  • Department of Interchangeable Manufacturing and Industrial Metrology, Vienna University of Technology, Vienna, Austria
References
  • 1. W. Shibo, G. Shirong, L. Hongtao, X. Huang, J. Biomimet. Biomater. Tissue Eng. 7, 7 (2010)
  • 2. H. Bhatt, T. Goswami, Biomed. Mater. 3, 042001 (2008)
  • 3. J.M. Martell, E. Berkson, R. Berger, J. Jacobs, J. Bone Joint Surgery Am. 85A, 1111 (2003)
  • 4. E.P.J. Watters, P.L. Spedding, J. Grimshaw, J.M. Duffy, R.L. Spedding, Chem. Eng. J. 112, 137 (2005)
  • 5. S. Ge, S. Wang, N. Gitis, M. Vinogradov, J. Xiao, Wear 264, 571 (2008)
  • 6. D. Xiong, S. Ge, Wear 250, 242 (2001)
  • 7. L. Mattei, F. DiPuccio, B. Piccigallo, E. Ciulli, Tribol. Int. 44, 532 (2011)
  • 8. T. Mizoue, K. Yamamoto, T. Masaoka, A. Imakiire, M. Akagi, I.C. Clarke, J. Orthoped. Sci. 8, 491 (2003)
  • 9. T. Masaoka, I.C. Clarke, K. Yamamoto, J. Tamura, P.A. Williams, V.D. Good, H. Shoji, A. Imakiire, Wear 254, 391 (2003)
  • 10. S. Carmignato, M. Spinelli, S. Affatato, E. Savio, Wear 270, 584 (2011)
  • 11. M.G. Teeter, D.D.R. Naudie, J.S. Milner, D.W. Holdsworth, J. Arthroplasty 26, 497 (2011)
  • 12. International Standard ISO 14242-2:2000, Implants for surgery - wear of total hip joint prostheses. Part 2. Methods of measurement
  • 13. A.E. Bowden, S.M. Kurtz, A.A. Edidin, J. Biomed. Mater. Res. Part B: Appl. Biomater. 75B, 205 (2005)
  • 14. K. Burckhardt, C. Gerber, J. Hodler, H. Notzli, G. Szekely, Med. Image Anal. 4, 375 (2000)
  • 15. P. Bills, L. Brown, X. Jiang, L. Blunt, J. Phys., Conf. Series 13, 316 (2005)
  • 16. L. Blunt, P. Bills, X. Jiang, C. Hardaker, G. Chakrabarty, Wear 266, 424 (2009)
  • 17. P. Bills, L. Blunt, X. Jiang, Wear 263, 1133 (2007)
  • 18. S.M. Kurtz, The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement, Elsevier Academic Press, London 2009, p. 511
  • 19. M.N. Durakbasa, P.H. Osanna, Measurement 33, 157 (2003)
  • 20. M.N. Durakbasa, P.H. Osanna, A. Afjehı-Sadat, D. Samarawıckrama, A. Krsek, Measur. Sci. Rev. 5(2), 1 (2005)
  • 21. M.N. Durakbasa, Geometrical Product Specifications and Verification for the Analytical Description of Technical and Non-Technical Structures, Abteilung Austauschbau und Masstechnik, Wien 2003, p. 126
  • 22. R.W.O.K. Rahmat, N.G. Seng Beng, K.A.P. Sangaralingam, Jurnal Teknologi 45(D), 97 (2006)
  • 23. O.K. Muratoglu, H.E. Rubash, C.R. Bragdon, B.R. Burroughs, A. Huang, W.H. Harris, J. Arthroplasty 22, 435 (2007)
  • 24. L.A. Blunt, P.J. Bills, X.Q. Jiang, G. Chakrabarty, Proc. Inst. Mech. Eng. H. 222, 309 (2008)
  • 25. D.M. Estok, II, C.R. Bragdon, G.R. Plank, A. Huang, O.K. Muratoglu, W.H. Harris, J. Arthroplasty 20, 239 (2005)
  • 26. O.K. Muratoglu, C.R. Bragdon, B.S, Murali Jasty, D.O. O'Connor, R.S. Von Knoch, W.H. Harris, J. Arthroplasty 19, 887 (2004)
  • 27. M. Spinelli, S. Carmignato, S. Affatato, M. Viceconti, Wear 267, 753 (2009)
  • 28. M.G. Teeter, D.D.R. Naudie, K.D. Charron, D.W. Holdsworth, J. Arthroplasty 25, 330 (2010)
  • 29. J.C. Grochowsky, L.W. Alaways, R. Siskey, E. Most, S.M. Kurtz, J. Biomed. Mater. Res. Part B: Appl. Biomater. 79, 263 (2006)
  • 30. R. Bogue, Sensor Rev. 30, 102 (2010)
  • 31. A. Akbari, M.E. Roy, L.A. Whiteside, B.J. Katerberg, D.J. Schnettgoecke, J. Arthroplasty, in press
Document Type
Publication order reference
Identifiers
YADDA identifier
bwmeta1.element.bwnjournal-article-appv121n139kz
JavaScript is turned off in your web browser. Turn it on to take full advantage of this site, then refresh the page.