2017 | 132 | 3 | 857-860
Article title

Determination of Pulse Width and Pulse Amplitude Characteristics of Materials Used in Pendulum Type Shock Calibration Device

Authors
Content
Title variants
Languages of publication
EN
Abstracts
EN
Calibration of shock and vibration transducers is performed by using the methods described in ISO 16063-13 and 16063-22 standards, at primary and secondary levels respectively. At secondary level, the devices, generating mechanical shock, are pendulum shock calibrator, dropball shock calibrator, pneumatically operated piston shock calibrator and Hopkinson bar shock calibrator. The main parameters and requirements of the shock calibrator devices, named here, are also given in ISO 16063-22 standard. Pulse width and pulse amplitude are the important parameters in calibration of the shock and vibration transducers. Pulse width and shape depend on the material properties. In this paper the determination of pulse width and pulse amplitude characteristics of materials used in pendulum type shock calibration device, designed in TUBİTAK UME, is presented.
Keywords
EN
Year
Volume
132
Issue
3
Pages
857-860
Physical description
Dates
published
2017-09
References
  • [1] ISO 16063-13, Methods for the calibration of vibration and shock transducers - Part 13: Primary shock calibration using laser interferometry, International Organization for Standardization, Geneva 2001
  • [2] ISO 16063-22, Methods for the calibration of vibration and shock transducers - Part 22: Shock calibration by comparison to a reference transducer, International Organization for Standardization, Geneva 2005
  • [3] A. Link, H.-J. von Martens, W. Wabinski, Proc. SPIE 3411 3411, 224 (1998), doi: 10.1117/12.307704
  • [4] H.-J. von Martens, Metrologia 36, 357 (1999), doi: 10.1088/0026-1394/36/4/16
  • [5] M. Peres, R.D. Sill, A New Solution for Shock and Vibration Calibration of Accelerometers in: http://www.modalshop.com, CAL LAB, 2007
  • [6] M. Kobusch, A. Link, A. Buss, T. Bruns, in: IMEKO 20th TC3, 3rd TC16 and 1st TC22 Int. Conf., Merida, Mexico 2007 http://www.imeko.org/publications/tc3-2007/IMEKO-TC3-2007-007u.pdf
  • [7] E. Bilgiç, Y. Durgut, Acta Phys. Pol. A 128, B-267 (2015), doi: 10.12693/APhysPolA.128.B-267
  • [8] H. Nozato, A. Oota, T. Ishigami, T. Usuda, Meas. Sci. Technol. 22, 125109 (2011), doi: 10.1088/0957-0233/22/12/125109
  • [9] Yu-Chung Huang, Jiun-Kai Chen, Hsin-Chia Ho, Chung-Sheng Tu, Chao-Jung Chen, Measurement 45, 2383 (2012), doi: 10.1016/j.measurement.2011.11.023
  • [10] H. Nozato, T. Bruns, H. Volkers, A. Oota, Meas. Sci. Technol. 25, 035005 (2014), doi: 10.1088/0957-0233/25/3/035005
  • [11] http://www.bipm.org, CCAUV.V-K4, Technical Protocol of CCAUV Key Comparison
Document Type
Publication order reference
YADDA identifier
bwmeta1.element.bwnjournal-article-app132z3-iip015kz
Identifiers
JavaScript is turned off in your web browser. Turn it on to take full advantage of this site, then refresh the page.