Full-text resources of PSJD and other databases are now available in the new Library of Science.
Visit https://bibliotekanauki.pl

PL EN


Preferences help
enabled [disable] Abstract
Number of results
2011 | 17 | 4 | 173-188

Article title

Analysis of the measurement uncertainty during calibration of radioherapy and radiodiagnostics dosemeters at the Polish Secondary Standard Dosimetry Laboratory

Content

Title variants

Languages of publication

EN

Abstracts

EN
The aim of the study was to work out a method of establishing the measurement uncertainty and determination of uncertainty values for calibration of dosemeters at the SSDL. For the determination of type A and B uncertainties, a method according to iternational ISO standard was used. The calibration procedures used at the SSDL, consistent with the reports of the International Atomic Energy Agency, were taken into account. The presented results apply to all dosimeters calibrated at the SSDL. The detailed analysis of every step of the calibration procedures and the results obtained show that the measurement uncertinty is within the limits accepted by the Polish Centre for Accreditation.
PL
Celem pracy było opracowanie metody szacowania niepewności i wstępne określenie wartości szacowanych wielkości w przypadku wzorcowanych w LWWD dawkomierzy. Zastosowano metodę obliczania niepewności typu A i B wg międzynarodowej normy ISO. Uwzględniono stosowany w LWWD sposób wzorcowania dawkomierzy wg raportów Międzynarodowej Agencji Energii Atomowej. Przedstawione wyniki dotyczą wszystkich wzorcowanych w LWWD dawkomierzy. Zarówno szczegółowa analiza wszystkich etapów wzorcowania jak i otrzymane wyniki wskazują, że niepewność pomiarów jest w zakresie akceptowanym przez Polskie Centrum Akredytacji.

Publisher

Year

Volume

17

Issue

4

Pages

173-188

Physical description

Dates

published
1 - 1 - 2011
online
26 - 4 - 2012

Contributors

  • Zakład Fizyki Medycznej, Centrum Onkologii, Warszawa, Polska
  • Zakład Fizyki Medycznej, Centrum Onkologii, Warszawa, Polska
  • Zakład Fizyki Medycznej, Centrum Onkologii, Warszawa, Polska

References

  • IAEA. Absorbed dose determination in photon and electron beams. An international code of practice. Technical Reports Series, Vienna, 1987; No TRS 277.
  • IAEA. Calibration of dosimeters used in radiotherapy. An international code of practice. Technical Reports Series, second edition, Vienna, 1994; No TRS 374.
  • IAEA. Absorbed dose determination in external beam radiotherapy. An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Technical Reports Series, Vienna, 2000; No TRS 398.
  • IAEA. Calibration of reference dosimeters for external beam radiotherapy. Technical Reports Series, Vienna, 2009; No TRS 469.
  • IAEA. Dosimetry in diagnostic radiology. An international code of practice. Technical Reports Series, Vienna, 2007; No TRS 457.
  • ISO/IEC/OIML/BIPM, Guide to the expression of uncertainty in measurements. 1995.